beautypg.com

Technical data, Designation value input, Sensor break function – KROHNE OPTITEMP TT 31 R EN User Manual

Page 22: Output, Ambient temperature, Galvanic isolation, Power supply, Typical accuracy, Installation

background image

02/2011 – 4000753501 – MA OPTITEMP TT31CR R01 en

www.krohne.com

22

6

Technical data

6.1 Specifications, TT 31 R / TT 31 R

6.1.1

Technical data

Designation

Value

Input

Pt100

-200...

+1000°C

Pt1000

-200...

+200°C

PtX

The upper range is dependent on

the X value.

Ni100

-60...

+250°C

Ni1000

-10...

+150°C

Potentiometer

0...

2000

Thermocouples

Types: AE, B, E, J, K, L, N, R, S, T, U

Voltage

-10...

+500 m

V

Sensor break function

User-defined output

Output

4...

20 mA, 20...4 mA,

temperature linear

Ambient temperature

TT 31 R

-20...

+70°C

TT 31 R

-20...

+60°C

Galvanic isolation

1500 VAC, 1 min

Power supply

TT 31 R

8...

36 VDC

TT 31 R

8...

36 VDC

Not intrinsically safe

Typical accuracy

+/- 0.1% of the measuring span

Installation

Rail according to EN 50020,

35 mm

Table 1: Technical Data TT 31 R / TT 31 R