Installation – KROHNE OPTIWAVE 5200 ATEX EN User Manual
Page 15
INSTALLATION
2
15
OPTIWAVE 5200 C/F
www.krohne.com
04/2013 - 4001905802 - AD ATEX OPTIWAVE 5200 R02 en
Remote (Field) versions (antenna housing only)
Equipment category 1/2 G or EPL Ga/Gb: Ex ia and Ex d ia devices
Temperature
class
Maximum ambient temperature
Max. flange
temperature
PP Wave Horn
PTFE Wave Horn
Wave-Stick
Metallic Horn
(standard
temperature)
Wave-Guide
Metallic Horn
(high-
temperature)
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
T6
+46
+115
+46
+115
+46
+115
+46
+115
+45
+113
T5
+41
+106
+41
+106
+41
+106
+44
+111
+55
+131
T4
+39
+102
+39
+102
+39
+102
+43
+109
+60
+140
Temperature
class
Minimum ambient temperature
Min. flange
temperature
PP Wave Horn
Wave-Stick
PTFE Wave Horn
Metallic Horn
(standard
temperature)
Wave-Guide
Metallic Horn
(high-
temperature)
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
T6...T2
-40
-40
-40
-40
-40
-40
-40
-40
-20
-4
INFORMATION!
Remote converter
Remote converter
Remote converter
Remote converter
The maximum ambient temperature, T
a
, permitted depends on the temperature class:
•
T
a
= +60
°
C for class T6
•
T
a
= +70
°
C for class T5
•
T
a
= +80
°
C for class T4
- BATCHFLUX 5500 C Quickstart EN (20 pages)
- IFC 050 Converter Quickstart EN (28 pages)
- IFC 100 Converter Quickstart EN (32 pages)
- IFC 300 Converter Quickstart EN (68 pages)
- OPTIFLUX 1000 Quickstart EN (20 pages)
- OPTIFLUX 2000 Quickstart EN (24 pages)
- OPTIFLUX 4000 Quickstart EN (24 pages)
- OPTIFLUX 4040C Quickstart EN (16 pages)
- OPTIFLUX 5000 Flange Quickstart EN (20 pages)
- OPTIFLUX 5000 Sandwich Quickstart EN (20 pages)
- OPTIFLUX 6000 Quickstart EN (28 pages)
- OPTIFLUX 7300 Quickstart EN (24 pages)
- OPTIPROBE Quickstart EN (16 pages)
- TIDALFLUX 2300 F EN (44 pages)
- TIDALFLUX 2300 F Quickstart EN (24 pages)
- WATERFLUX 3000 EN (40 pages)
- WATERFLUX 3000 Quickstart EN (24 pages)
- WATERFLUX 3070 EN (80 pages)
- WATERFLUX 3070 Quickstart EN (32 pages)
- USB ADAPTER PLUS EMF EN (16 pages)
- IFC 050 Converter Modbus EN (20 pages)
- IFC 100 Converter FOUNDATION FIELDBUS EN (64 pages)
- IFC 100 Converter Modbus EN (20 pages)
- IFC 300 Converter FOUNDATION FIELDBUS EN (60 pages)
- IFC 300 Converter HART 0102 EN (20 pages)
- IFC 300 Converter HART 0201 EN (23 pages)
- IFC 300 Converter Modbus EN (24 pages)
- IFC 300 Converter PROFIBUS PA DP EN (40 pages)
- OPTIFLUX 2000-4000 IECEx EN (16 pages)
- OPTIFLUX 2000-4000-5000-6000-7300-IFC 300 Ex EN (40 pages)
- OPTIFLUX 2000-4000-5000-6000 -IFC 100 Ex EN (24 pages)
- OPTIFLUX 4040 C Ex EN (20 pages)
- OPTIFLUX x300 Ex Zone2 EN (1 page)
- H250 M9 ES EN (36 pages)
- VA 40-VA 45 EN (36 pages)
- H250 M10 ATEX II2G Ex d EN (16 pages)
- H250 M10 ATEX II3D Ex t EN (16 pages)
- H250 M40 ATEX II2D Ex t-II2G Ex d EN (20 pages)
- H250 M40 ATEX II2G Ex i EN (20 pages)
- H250 M40 ATEX II3G Ex nA EN (20 pages)
- H250 M40 Ex II2G Reed EN (2 pages)
- H250 M9 ATEX II2G Ex i EN (16 pages)
- H250 M9S ATEX II3D Ex t-II3G Ex nA EN (20 pages)
- M8E Converter HART 0101 EN (13 pages)
- DK 32-DK 34 ATEX II2G Ex i EN (16 pages)