General safety information – KROHNE OPTIWAVE 5200 ATEX EN User Manual
Page 10
1
GENERAL SAFETY INFORMATION
10
OPTIWAVE 5200 C/F
www.krohne.com
04/2013 - 4001905802 - AD ATEX OPTIWAVE 5200 R02 en
Figure 1-8: Remote version, Ex ic nameplate on the converter housing
Figure 1-9: Remote version, Ex ic nameplate on the antenna housing
1 ATEX certification agency code
2 Equipment approval category (explosive atmosphere - gas), types of device protection including approved Gas Groups,
temperature classes (T6...T4 or T3 or T2 - depends on the antenna type) and equipment protection level
3 Equipment category (explosive atmosphere - dust), types of device protection for explosive atmospheres with dust,
zones, maximum surface temperature and degree of ingress protection (if fitted with the appropriate cable glands)
4 4...20 mA passive - HART output option:
4...20 mA passive - HART output option:
4...20 mA passive - HART output option:
4...20 mA passive - HART output option: Intrinsically-safe circuit data
Fieldbus (FF or PROFIBUS PA) options:
Fieldbus (FF or PROFIBUS PA) options:
Fieldbus (FF or PROFIBUS PA) options:
Fieldbus (FF or PROFIBUS PA) options: Entity or FISCO power supply parameters
5 Input voltage range and maximum current (4...20 mA passive - HART) / basic current (FF or PROFIBUS PA)
6 Cable entry type and size (M20×1.5 or ½ NPT)