TECHNICAL DATA

24

OPTISWIRL 4070

www.krohne.com

06/2010 - 4000122102 - QS OPTISWIRL R04 en

Technical data

4.1 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local representative.

Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Application range

Flow measurement of liquids, gases and vapours

Operating method / measuring principle

Karman vortex street

Measured value

Measured value

Measured value

Measured value

Primary measured value

Number of separated vortices

Secondary measured value

Operating and standard volumetric flow, mass flow

Signal converter

Signal converter

Signal converter

Signal converter

Versions

Compact

Option: Ex version

Measuring sensor

Measuring sensor

Measuring sensor

Measuring sensor

Standard

Basic device in flange version (with integrated temperature measurement)

Basic device in sandwich version (with integrated temperature

measurement)

Optional

Basic device with additional pressure measurement

Basic device with additional pressure measurement and shut-off valve for

pressure sensor

Dual measuring device in both flange and sandwich version (redundant

measurement)

Dual measuring device in flange version with additional pressure

measurement

Display and user interface

Display and user interface

Display and user interface

Display and user interface

Local display

2 lines, 10 characters per line

Operating and display languages

German, English, French

Measuring accuracy

Reference condition

Water at 20ºC

Air at 20°C and 1.013 bar abs.

Accuracy

Based on volume flow

Liquids

±

0.75% of measured value (Re ≥ 20000)

±

2.0% of measured value (10000 < Re < 20000)

Gases and steams

±

1.0% of measured value (Re ≥ 20000)

±

2.0% of measured value (10000 < Re < 20000)

Pressure and temperature compensation: ±1.5% of measured value (Re ≥

20000); ±2.5% of measured value (10000 < Re < 20000)

Repeatability

±

0.1% of measured value

Long-term stability

±

0.01% of measured value

QS_OPTISWIRL4070_R04_en_122102_PRT.book Page 24 Monday, June 7, 2010 8:51 AM