TECHNICAL DATA
7
35
VA40 - VA45
www.krohne.com
09/2013 - 4000424402 MA VA40 45 R04 en
Measuring ranges VA45
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:
Reference condition for gas measurements:
The flow measurement of gases are refered to
Nl/h or Nm
3
/h: Volume flow in Normal state 0°C, 1.013 bar abs. (DIN 1343)
SCFM or SCFH: Volume flow in Standard state 15°C, 1.013 bar abs. (ISO 13443)
Measuring span:
10 : 1
Declaration of flow::
Values = 100%
Air 20°C [68°F], 1.013 bar abs. [14,7 psia]
1 float form C
2 float form D
3 reading line
Cone no.
Float
form
Flow air
Max. pressure loss
[l/h]
[SCFH]
[mbar]
[psi]
DN15
N -15.01
C
1500...2300
55,8...85,6
3
0,044
D
2300…4800
85,6...179
3
0,044
N -15.02
C
5500…9000
205...335
3
0,044
D
9000…16000
335...595
3
0,044
DN25
N -25.01
C
3000…5000
112...186
3
0,044
D
5000…7500
186...279
3
0,044
N -25.02
C
7500…16500
279...614
3
0,044
D
16500…25000
614...930
4
0,058
DN40
N -40.01
C
17000…26000
632...967
4
0,058
D
26000…34000
967...1265
4
0,058
N -40.02
C
34000…60000
1265...2232
4
0,058
D
60000…75000
2232...2790
4
0,058
INFORMATION!
The operating pressure should be at least 5 times the pressure loss for gases! The specified
pressure drops are valid for water and air at maximum flow rate. Other flow ranges on request.
Conversion of other media or operating data (pressure, temperature, density, viscosity) is
performed using the calculation method in accordance with VDI /VDE Directive 3513