Modbus protocol – KROHNE IFC 100 Converter Modbus EN User Manual
Page 19
MODBUS PROTOCOL
6
19
IFC 100
www.krohne.com
08/2014 - 4001410101 - AD Modbus IFC 100 R01 en
6.10.3 Conductivity Calibration
1
User enters the actual conductivity in S/m
User enters the actual conductivity in S/m
User enters the actual conductivity in S/m
User enters the actual conductivity in S/m
This is the target for the calibration
2
Write actual conductivity
Write actual conductivity
Write actual conductivity
Write actual conductivity
Write Holding Register 43034 type float
(Modbus Fct. 0x10)
3
Write actual conductivity
Write actual conductivity
Write actual conductivity
Write actual conductivity
Write Holding Register 43036 type float
(Modbus Fct. 0x10)
3a
The target value has to be written to both
registers!
4
Start conductivity calibration:
Start conductivity calibration:
Start conductivity calibration:
Start conductivity calibration:
Set single coil 2002 to 1 (Modbus Fct. 0x05)
5
Check status of calibration:
Check status of calibration:
Check status of calibration:
Check status of calibration:
Read single coil 2002 (Modbus Fct. 0x01)
5a
[2002 == 1: calibration running]
5b
[2002 == 0: calibration complete]
6
Read Calibration Value
Read Calibration Value
Read Calibration Value
Read Calibration Value
Read Input Register 20000 type float
(Modbus Fct. 0x04)
7
Write the calibration value
Write the calibration value
Write the calibration value
Write the calibration value
Write Holding Register 43032 type float
(Modbus Fct. 0x10)
8
Activate new values
Activate new values
Activate new values
Activate new values
Set single coil 1001 to 1 (Modbus Fct. 0x05)
8a
This results in a warm start of the device.
Will take maximum 5s.
.book Page 19 Monday, July 28, 2014 11:32 AM