Modbus protocol, 10 calibration procedures – KROHNE IFC 100 Converter Modbus EN User Manual
Page 17
MODBUS PROTOCOL
6
17
IFC 100
www.krohne.com
08/2014 - 4001410101 - AD Modbus IFC 100 R01 en
6.10 Calibration procedures
6.10.1 Zero Flow Calibration
1
User reduces the actual flow through the meter to
User reduces the actual flow through the meter to
User reduces the actual flow through the meter to
User reduces the actual flow through the meter to
zero
zero
zero
zero
2
Start zero calibration:
Start zero calibration:
Start zero calibration:
Start zero calibration:
Set single coil 2000 to 1 (Modbus Fct. 0x05)
3
Check status of calibration
Check status of calibration
Check status of calibration
Check status of calibration
Read single coil 2000 (Modbus Fct. 0x01)
3a
[2000 == 1: calibration running]
3b
[2000 == 0: calibration complete]
4
Read Calibration Value
Read Calibration Value
Read Calibration Value
Read Calibration Value
Read Input Register 20000 type float
(Modbus Fct. 0x04)
5
User checks the calibration value:
User checks the calibration value:
User checks the calibration value:
User checks the calibration value:
-0.01...+0.01: good results
-0.1...+0.1: acceptable result for difficult
application
5a
Not ok
5b
ok
6
User tries to improve the calibration situation
User tries to improve the calibration situation
User tries to improve the calibration situation
User tries to improve the calibration situation
6a
Improvement done
6b
Improvement not possible
-Break-
7
Write the calibration value
Write the calibration value
Write the calibration value
Write the calibration value
Write Holding Register 43024 type float
(Modbus Fct. 0x10)
8
Activate new values
Activate new values
Activate new values
Activate new values
Set single coil 1001 to 1 (Modbus Fct. 0x05)
8a
This results in a warm start of the device.
Will take maximum 5s.
.book Page 17 Monday, July 28, 2014 11:32 AM