Output characteristics, Environmental characteristics, General characteristics – KEPCO JBW 75W Series User Manual
Page 2

2
228-1551 REV 2
061510
KEPCO, INC. 131-38 SANFORD AVENUE FLUSHING, NY. 11355 U.S.A. TEL (718) 461-7000 FAX (718) 767-1102
http://www.kepcopower.com email: [email protected]
LEAKAGE CURRENT: Cold start:
100V a-c and 60 Hz (single pole switching) (operating
in conformance with Den-An):
0.75mA max, 0.2mA typ
240V a-c and 60 Hz (single pole switching in confor-
mance to UL 1950/IEC 950):
0.75mA max, 0.3mA typ
POWER FACTOR:
100V a-c: 0.99 typ.
240V a-c: 0.95 typ.
OUTPUT CHARACTERISTICS:
SOURCE EFFECT: (85 to 265 Va-c)
0.4% max.
LOAD EFFECT: measured at sensing terminals
(0% - 100% load change)
0.8% max.
TEMPERATURE EFFECT: (-10 to 60°C)
1.0% max.
COMBINED EFFECT:
2.0% max. (4% typical for overshoot at start-up)
TIME EFFECT OR DRIFT: (1/2 to 8 hr. at 25°C)
0.4% max.
TRANSIENT RECOVERY: A step load change from
50% to 100% of rated load in 50 microseconds or more
produces no more than 4% output voltage excursion.
Temperature range from -10 to 60°C. Recovery time is
1mS maximum.
HOLD UP TIME: Condition: 100% output load. Upon
input interruption the output is maintained for:
100 Va-c: 40mS typ.
240 Va-c: 50mS typ.
START UP TIME:
100 Va-c: 500mS max., 400mS typ.
240 Va-c: 350mS max., 300mS typ.
OVERVOLTAGE PROTECTION: Fixed, factory set.
See Table 1. Latching will occur, reset time: 1min. max.
OVERCURRENT: Square type, output voltage returns
to rated level upon removal of cause of malfunction (long
term overcurrent could damage unit).
ENVIRONMENTAL CHARACTERISTICS
OPERATING TEMPERATURE: -10 to 60°C (start up
-20 to -10°C). See the derating, Figure 1.
STORAGE TEMPERATURE: -30°C to + 75°C.
COOLING: Natural convection. Do not allow the power
supply to become dust covered because that will
decrease the cooling efficiency of the unit and cause insu-
lation to deteriorate.
ORIENTATION: Vertical or horizontal (see Figure 2).
HUMIDITY: 10% to 90% relative humidity, operating
and storage, noncondensing, wet bulb temperature < or =
35°C.
VIBRATION: Three axes, one hour each, sweep time
10 min., nonoperating.
5-10 Hz., 10 mm amplitude
10-200 Hz., 2G (19.6m/s
2
) acceleration
SHOCK: Three axes, 60G (588m/s
2
), 11 mS ±5mSec
pulse duration, three shocks each axis, nonoperating, 1/2
sine pulse.
GENERAL CHARACTERISTICS
WITHSTANDING VOLTAGE: (at 5 to 35°C ambient,
45 to 85% relative humidity, cutout current 10 mA):
Between input and output terminals:
3.0 KV a-c for 1 minute.
Between input terminals and ground:
2000V a-c for 1 minute.
Between output terminals and ground:
500V a-c for 1 minute.
INSULATION RESISTANCE: Between input and out-
put, input and ground, output and ground: 100 megohms
minimum (500V d-c, 5 to 35°C ambient, 45 to 85% relative
humidity)
SAFETY: All units designed to meet UL 60950, CSA
C22.2,No. 60950-1(C-UL), and TÜV Rheinland EN60950
(ambient temp. 50°C). JBW 75W units are CE marked per
the Low Voltage Directive (LVD), EN60950.
EMC - EMISSIONS:
• Conducted Noise 0.15MHz to 30MHz:
FCC Class B, VCCI-B, EN55011-B, EN55022-B.
• Input Harmonics (on AC Mains) 0 to 2kHz:
EN 61000-3-2.
EMC - IMMUNITY: Designed to meet EN 61000-6-2.
• ESD:
EN 61000-4-2 Level 4, Normal operation.
• Radiated Field Noise:
EN 61000-4-3 Level 3, Normal operation.
• Electrical Fast Transient/Burst (EFT):
EN 61000-4-4 Level 3, Normal operation.
• Surge:
EN 61000-4-5 Level 4, no damage.
• Conducted Noise:
EN 61000-4-6 Level 3, Normal operation.
• Power Frequency Magnetic Field:
EN 61000-4-8 Level 4, Normal operation.
• Interruptions and voltage dips, short variations:
EN 61000-4-11, Normal operation.
WARRANTY: One year.