6.2. GENERAL
CHARACTERISTICS
6.2.1. Reference
standards
Safety: IEC/EN
61010-1
Insulation:
double insulation
Pollution level:
2
Measurement category:
CAT I 30V
Max operating altitude:
2000m
6.2.2. General
characteristics
Mechanical characteristics
Size (L x W x H):
195 x 92 x 55mm
Weight (battery included):
400g
Display
Characteristics:
5 LCD, decimal sign and point
Over range indication:
the display shows message “-OL-”
Power supply
Rechargeable battery
1x7.4/8.4V 700mAh Li-ION
Alkaline battery:
1x9V type NEDA1604 006P IEC6F22
External adapter:
230VAC/50Hz – 12VDC/1A
Battery life:
SOUR mode: approx. 8 hours (@ 12mA, 500
)
MEAS/SIMU mode: approx. 15 hours
Low battery indication:
the display shows symbol “ ”
Auto power off:
after 20 minutes (adjustable) of non-operation
6.3. ENVIRONMENT
6.3.1. Environmental conditions for use
Reference temperature:
18°C
28°C
Operating temperature:
-10 ÷ 40°C
Allowable relative humidity:
<95%RH up to 30°C, <75%RH up to 40°C
<45%RH up to 50°C, <35%RH up to 55°C
Storage temperature:
-20 ÷ 60°C
This instrument satisfies the requirements of Low Voltage Directive 2006/95/EC
(LVD) and of EMC Directive 2004/108/EC
6.4. ACCESSORIES
6.4.1. Accessories
provided
Pair of test leads
Pair of alligator clips
Protection
shell
Rechargeable battery (not inserted)
External battery charger
User
manual
Hard carrying case