4.3.5 Functional test of the receiver
If problems or strange measuring results occur during a line location procedure, it may be
helpful to check the functionality of the receiver using the integrated self-test. This test may
provide information about possible malfunctions of the instrument. Preferably, the self-test
should be performed in an interference-free environment without live cables
Proceed as follows to perform a self-test:
Step Action
1
Switch off the receiver RX5000. The test can be started by pressing
and holding
followed by
for at least 3 seconds
If the self-test is successfully accessed, the display should show the
software and hardware versions of the receiver components:
2
Press the
function key to start the self-test
3
If the test is successfully passed without any problems, the display
shows the following information:
If a fault is recognized during the test, the self-test procedure is
terminated and the display indicates the respective error code (see
next page):
If this is the case, the error code should be forwarded to the service
department of HT ITALIA
4
Press the
function key to exit the self-test