Menu Flow Chart

Figure 9 Menu Flow Chart

6.4.1 Set # of Channels
This menu entry allows the selection of the number of active channels. NOTE: the number of activated
channels can be less than (but not greater than) the controller’s maximum input capacity. It should match
the number of sensors you are hooking up to the controller. The menu appears as:

6.4.2 Set Channel Range

NORMAL
OPERATION

MAIN MENU:

SET # OF CHANNELS

SET # OF CHANNELS
X

MAIN MENU:

SET CHANNEL RANGE

SET CHANNEL RANGE
CHX = #####

CH1 … X

MAIN MENU:

SET CHANNEL TYPE

SET CHANNEL TYPE
CHX = XXXXXXX

CH1 … X

MAIN MENU:

SET CHANNEL ALARMS

SET CHANNEL
ALARMS

CH1 … X

MAIN MENU:
SET DATE & TIME

SET DATE AND TIME
MM/DD/YY HH:MM:SS

MAIN MENU:

VIEW TWA & PEAK

01/02/04

1200 - 1300

TWA #### PK #####

MAIN MENU:

VIEW ALARM RECS

ALM: MM/DD/YY
CHX- A1 ON HH:MM:SS

KEY:

Up Arrow

Down Arrow

Enter

PROG

Set # of Channels:

1

This menu entry allows the selection of the Channel Range for each gas channel input. This is a scrolling
list from 1.00 up to 10,000 and covers all of the Detcon gas sensor range possibilities. This menu appears
as:

Set Channel Range:
CH1 = XX.X

NOTE: If the range is changed, the alarm levels will have to be reset!