5. Service and Maintenance
Calibration Frequency
In most applications, monthly to quarterly span calibration intervals will assure reliable detection. Upon initial
installation and commissioning, close frequency tests should be performed, weekly to monthly. Test results
should be recorded and reviewed to determine a suitable calibration interval due to differences in industrial
environments.
Visual Inspection
The Sensor should be inspected annually for the following:
• Inspect the sensor for signs of corrosion, pitting, and water damage.
• Remove the Splash Guard and inspected it for blockage, broken, cracked, or missing pieces.
• For H2S Sensor assemblies, inspect CXT-IR Series Splashguard Adapter Assembly with integral filter
(PN 602-003552-100) for blockage of filter material.
• Inspect inside of the Junction Box for signs of water accumulation, signs of corrosion.
• Check wiring to ensure there are no loose or pinched wires and all connections are clean and tight.
Condensation Prevention Packet
A moisture condensation packet (Detcon PN 960-202200-000) should be installed in every explosion proof
Junction Box. The moisture condensation prevention packet will prevent the internal volume of the J-Box from
condensing and accumulating moisture due to day-night humidity changes. This packet provides a critical
function and should be replaced annually.
5.1 Replacement of the Batteries/Battery Pack
5.1.1 Units with 12V Smart Battery Pack
The CXT-IR Sensor incorporates a ‘Low battery’ fault message which will appear as an alternating display
between the current gas reading and ‘VOLT’ on the sensor display. This fault will appear when the battery
voltage drops below 7.5 volts. When the voltages drops below 7.2 volts the display will change to a steady
‘VOLT’ display, the sensor will then enter into ‘Fault’ mode, and the sensor will no longer provide a valid gas
reading.
NOTE
The safety approvals require removing entire sensor assembly to a non-hazardous area
before changing out the batteries or battery pack.
1. Remove the cover from the J-box.
2. Unlatch the holding lever on the side of the battery bracket, and swing the cover away from the battery
pack.
3. Remove the battery pack from the J-box, and replace with new battery pack.
4. Swing the cover back over the battery pack, and ensure that it is latched.
5. Replace the cover on the J-box.
5.1.2 Units with Tri-‘C’ sized Lithium Battery Holder
NOTE
The safety approvals require removing entire sensor assembly to a non-hazardous area
before changing out the batteries or battery pack.