Model FP-624D

FP-624D Instruction Manual

Rev. 1.2

Page 40 of 42

1000 ohms maximum loop load @ 24VDC
0mA

All Fault Diagnostics

2mA

In-Calibration

4-20mA

0-100% full-scale

22mA

Over-range condition

Serial Output:

RS-485 Modbus™ RTU
Baud Rate 9600 BPS (9600, N, 8, 1 Half Duplex)

Relay Outputs:

Alarm 1, Alarm 2, and Fault
5A @ 250VAC
5A @ 30VDC

Status Indicators:

4-digit LED Display with gas concentration, full-script menu
Prompts for AutoSpan, AutoZero, Set-up Options, and Fault Reporting

Faults Monitored:

Loop Fault, Input Voltage Fault, Zero Fault, Sensor Fault,
Processor Fault, Memory Fault, Calibration Fault(s)
4 LEDs for Alarm 1, Alarm 2, Fault, and Calibration

Cable Requirements:

Power/Analog: 3-wire shielded cable
Maximum distance is 13,300 feet with 14 AWG

Serial Output: 2-wire twisted-pair shielded cable specified for RS-485 use
Maximum distance is 4,000 feet to last sensor

Mechanical Specifications

Length:

8 inches (200mm), including Splashguard

Width:

5.5 inches (140mm)

Weight:

2.7lbs (3.4Kg)

Mechanical Connection:

¾” Male NPT threaded connection

Electrical Connection:

Eleven 14 gauge (maximum) wire terminal landings