Once a new material/specie is selected, control returns to the Create Job

screen.

Click SELECT on the ALARM entry to configure meter’s set-point values:

The “ALARM” feature in the meter provides the user with a set-point:

readings above a specified value (considered “WET” or unacceptable)

generate an audible alarm.

The first reading breaking the alarm set-point will be signaled by four short

beeps in quick sequence. Subsequent readings that remain above the set-

point will be signaled by a single short beep. A reading below the

threshold will cause the alarm signal to reset, such that any following

reading above the threshold will be once again signaled by four beeps.

Once ALARM is chosen, pressing the SELECT key toggles the state of

the alarm feature (ON-OFF). The ON state is marked by the current alarm

threshold value, while the OFF state is indicated with an “OFF” statement.

When the alarm is ON, pressing the LEFT or RIGHT keys will change the

threshold value (down and up, respectively). Threshold values range from

0.05% MC to 59.5% MC (default is 8%).

Select “DONE” to return to the Create Job menu.

Selecting “DONE” in the Create Job menu returns the user to the MC

Read screen. The meter is now configured with the new job and all

parameters defined in the “Create Job” screen. Selecting “CANCEL” in the

Create Job menu returns the user to the Job Select screen; where it has

the opportunity to re-start the job select/create process. Pressing CANCEL

in the Job Select screen brings the user back to MC READ. The meter

maintains the configuration it had before the JOB option was selected.