Once a new material/specie is selected, control returns to the Create Job
screen.
Click SELECT on the ALARM entry to configure meter’s set-point values:
The “ALARM” feature in the meter provides the user with a set-point:
readings above a specified value (considered “WET” or unacceptable)
generate an audible alarm.
The first reading breaking the alarm set-point will be signaled by four short
beeps in quick sequence. Subsequent readings that remain above the set-
point will be signaled by a single short beep. A reading below the
threshold will cause the alarm signal to reset, such that any following
reading above the threshold will be once again signaled by four beeps.
Once ALARM is chosen, pressing the SELECT key toggles the state of
the alarm feature (ON-OFF). The ON state is marked by the current alarm
threshold value, while the OFF state is indicated with an “OFF” statement.
When the alarm is ON, pressing the LEFT or RIGHT keys will change the
threshold value (down and up, respectively). Threshold values range from
0.05% MC to 59.5% MC (default is 8%).
Select “DONE” to return to the Create Job menu.
Selecting “DONE” in the Create Job menu returns the user to the MC
Read screen. The meter is now configured with the new job and all
parameters defined in the “Create Job” screen. Selecting “CANCEL” in the
Create Job menu returns the user to the Job Select screen; where it has
the opportunity to re-start the job select/create process. Pressing CANCEL
in the Job Select screen brings the user back to MC READ. The meter
maintains the configuration it had before the JOB option was selected.