Hygrocontrol manual for type 83 page - 8 – COSA Xentaur Hygrocontrol Type 83 User Manual
Page 9
HYGROCONTROL Manual for Type 83 Page
-
8
-
5
.2.6.
List of
Commands
:
First an "F" for FullPromt output must be sent.
Each command must be ended by
Hxxx : HUMID.CAL – starts a calibration with the humidity described by xxx,
where xxx="345" means 34,5% RH
D : DUMP - sends the values in the calibration memory
E : ERASE - deletes the calibration memory
F : FullPromt - switches over to "FullPromt" output
Txxxx : calibration of temperature with xxxx="2345" means 23,45°C
Y : YES - positive confirmation of request
N : NO - negative confirmation of request
S : ESC - cancellation before normal end of calibration
5.2.7.
Output Examples:
Normal outputs of measured values:
R=43988 F=47447 Q=22709 H=5558 T=2740 DAC(h) = 8E48
ADC(h) = 987D
After the start of a calibration with „HUMID.CAL“:
CAL 0: R=43988 F=47456 Q=22962 H=5526 STOP? => S
If calibration was ended successfully:
CAL 0: R=43988 F=47457 Q=22962 H=5526 OK? Y/N
The answer to a command CALIB.TAB will be:
ADC calib val: 44793
Temp Offset : -19
Error Profile : 1
[ 0] Conc:= 0 Ratio:=19379
[ 1] Conc:= 100 Ratio:=20111
[ 2] Conc:= 200 Ratio:=20725
[ 3] Conc:= 350 Ratio:=21608
[ 4] Conc:= -1 Ratio:=65535
[ 5] Conc:= 500 Ratio:=22408
[ 6] Conc:= 650 Ratio:=23217
[ 7] Conc:= 800 Ratio:=24071
[ 8] Conc:= -1 Ratio:=65535
[ 9] Conc:= 950 Ratio:=24946
[10] Conc:= 0 Ratio:=19379
[11] Conc:= 800 Ratio:=24071
press RET.
The last 2 values [10] and [11] are the so-called „initial“ calibration
values done by Hygrocontrol, which cannot be deleted by customers.