Hygrocontrol manual for type 83 page - 9 – COSA Xentaur Hygrocontrol Type 83 User Manual
Page 10
HYGROCONTROL Manual for Type 83 Page
-
9
-
Press RET will jump back to normal measurement mode..
ERASE: this command will delete the calibration values [ 0] to [ 9] if
You confirm the request OK? [Y/N] by answering Y.
Tempcal:
Temperature Offset -20 is in the memory! (=correction by -0,2°C).
FullPromt:
R=43988 F=47456 Q=22962 H=5526 T=26570 DAC(h) = 8E48
ADC(h) = 987D
5.2.8
Used abbreviations
are
:
R:
Digits of Reference Oscillator
F:
Digits of Humidity Oscillator
Q:
Quotient Reference/Humidity
H:
Humidity in 1/100 %RH.
T:
Temperature in 1/100 °C
DAC(h):
Digital value of Analogue-Digital-Converter
(hexadecimal)
DAC(h):
Digital value of Digital-Analogue-Converter
(hexadecimal)
ADC calib val: Digital value of Analogue-Digital-Converter at 70°C
Temp Offset: Difference between measured and real Temperature
in 1/100 °C
Error Profile: x=1 output<4mA; x=2 hold output; x=3 output > 20mA
Conc:
Point of calibration in 1/10 °C
Ratio:
Quotient at calibration point