HYGROCONTROL Manual for Type 83 Page

-

9

-

Press RET will jump back to normal measurement mode..

ERASE: this command will delete the calibration values [ 0] to [ 9] if
You confirm the request OK? [Y/N] by answering Y.

Tempcal:
Temperature Offset -20 is in the memory! (=correction by -0,2°C).

FullPromt:
R=43988 F=47456 Q=22962 H=5526 T=26570 DAC(h) = 8E48

ADC(h) = 987D

5.2.8

Used abbreviations

are

:

R:

Digits of Reference Oscillator

F:

Digits of Humidity Oscillator

Q:

Quotient Reference/Humidity

H:

Humidity in 1/100 %RH.

T:

Temperature in 1/100 °C

DAC(h):

Digital value of Analogue-Digital-Converter
(hexadecimal)

DAC(h):

Digital value of Digital-Analogue-Converter
(hexadecimal)

ADC calib val: Digital value of Analogue-Digital-Converter at 70°C

Temp Offset: Difference between measured and real Temperature

in 1/100 °C

Error Profile: x=1 output<4mA; x=2 hold output; x=3 output > 20mA
Conc:

Point of calibration in 1/10 °C

Ratio:

Quotient at calibration point