Hygrocontrol manual for type 83 page - 3 – COSA Xentaur Hygrocontrol Type 83 User Manual
Page 4
HYGROCONTROL Manual for Type 83 Page
-
3
-
Please take care that cable diameter corresponds with the dimension
of the gland and so will guarantee a dense screwing. During
operation the housing of the instrument should be closed.
2.3 Instructions for Dispatch
If there any problems you are not able to correct, we ask you to
contact our technical department or to return the instrument with a
short description of the fault. When returning please do not forget to
pack the instrument safely in a cardboard box and fill up holes with
soft packing material to avoid damage.
3.
Technical Data
3.1 Humidity
Humidity Range
0....100 % RH.
Resolution
0,1 % RH.
Humidity Sensor
SE - 02
Temperature Error of Humidity
± 0,01 % RH./ °C
Repeatability
< 0,5 % RH.
Hysteresis after 4 hour Cycle
10 % - 95 % - 10 % r. F.
< 0,5 % RH.
Linearity of Electronic / Humidity
< 1 % RH.
3.2 Temperature
Range
-50.....+150 °C
Resolution
0,01 °C
Temperature Sensor
RTD: Pt-1000
System Accuracy for Temperature
over the total Range
± 0,35 °C
Linearity of Electronic / Temperature
± 0,15 °C
3.4 Maximum environmental conditions during operation
Transmitter Electronics
-10...+ 50 °C
Probe Tube with Sensors
-50...+150 °C
Max. Humidity
95%RH
3
.5
Connections
Screw Terminals
Marked with + and –