Please take care that cable diameter corresponds with the dimension
of the gland and so will guarantee a dense screwing. During
operation the housing of the instrument should be closed.

2.3 Instructions for Dispatch

If there any problems you are not able to correct, we ask you to
contact our technical department or to return the instrument with a
short description of the fault. When returning please do not forget to
pack the instrument safely in a cardboard box and fill up holes with
soft packing material to avoid damage.

3.

Technical Data

3.1 Humidity

Humidity Range

0....100 % RH.

Resolution

0,1 % RH.

Humidity Sensor

SE - 02

Temperature Error of Humidity

± 0,01 % RH./ °C

Repeatability

< 0,5 % RH.

Hysteresis after 4 hour Cycle
10 % - 95 % - 10 % r. F.

< 0,5 % RH.

Linearity of Electronic / Humidity

< 1 % RH.

3.2 Temperature

Range

-50.....+150 °C

Resolution

0,01 °C

Temperature Sensor

RTD: Pt-1000

System Accuracy for Temperature
over the total Range

± 0,35 °C

Linearity of Electronic / Temperature

± 0,15 °C

3.4 Maximum environmental conditions during operation

Transmitter Electronics

-10...+ 50 °C

Probe Tube with Sensors

-50...+150 °C

Max. Humidity

95%RH

3

.5

Connections

Screw Terminals
Marked with + and –