HYGROCONTROL Manual for Type 83 Page
-
12
-
results of a measurement of saturated vapour over ice at different
temperatures below zero.
Table 3: Display at 100% RH over Ice at different Temperatures
Temperature (°C)
PSF (mbar)
PSE (mbar)
M (%RH)
0
6,11
6,11
100,0
- 5
4,22
4,02
95,3
-10
2,87
2,60
90,6
-15
1,91
1,66
86,9
-20
1,26
1,03
81,7
-25
0,81
0,64
79,0
-30
0,49
0,37
75,5
PSF = Saturated Vapour Pressure over Chilled Water
PSE = Saturated Vapour Pressure over Ice
M
= Relative Humidity displayed by HYGROMESS Instruments.
6.7 Measurements under Pressure or Vacuum
All probes can be used at pressures of 0,9 to 1,3 bar. Only special
probes with a glass feed-through (which predicts SS - 316 tube
material) may be used under pressures of 0,03 to 30 bar. As relative
Humidity is directly proportional to the pressure (for example: increase
of pressure by factor 2 also increases relative Humidity by factor 2,
when temperature and content of water remain constant!) you should
take care of a place with constant pressure values, when mounting
the probe. Be aware that pressure may change, when diameters of
tubes change - and air streams may be turbulent at these places.