beautypg.com

Hygrocontrol manual for type 83 page - 12 – COSA Xentaur Hygrocontrol Type 83 User Manual

Page 13

background image

HYGROCONTROL Manual for Type 83 Page

-

12

-

results of a measurement of saturated vapour over ice at different
temperatures below zero.

Table 3: Display at 100% RH over Ice at different Temperatures

Temperature (°C)

PSF (mbar)

PSE (mbar)

M (%RH)

0

6,11

6,11

100,0

- 5

4,22

4,02

95,3

-10

2,87

2,60

90,6

-15

1,91

1,66

86,9

-20

1,26

1,03

81,7

-25

0,81

0,64

79,0

-30

0,49

0,37

75,5

PSF = Saturated Vapour Pressure over Chilled Water
PSE = Saturated Vapour Pressure over Ice
M

= Relative Humidity displayed by HYGROMESS Instruments.

6.7 Measurements under Pressure or Vacuum

All probes can be used at pressures of 0,9 to 1,3 bar. Only special
probes with a glass feed-through (which predicts SS - 316 tube
material) may be used under pressures of 0,03 to 30 bar. As relative
Humidity is directly proportional to the pressure (for example: increase
of pressure by factor 2 also increases relative Humidity by factor 2,
when temperature and content of water remain constant!) you should
take care of a place with constant pressure values, when mounting
the probe. Be aware that pressure may change, when diameters of
tubes change - and air streams may be turbulent at these places.