Troubleshooting – Seca 645 User Manual
Page 109

Troubleshooting • 109
If you find faults or deviations during the function
check, as a first step try to remedy the fault with the aid
of the section entitled "Troubleshooting" from
page 109.
CAUTION!
Bodily injury
If you find faults or deviations during the func-
tion check which you are unable to remedy with
the aid of the section entitled "Troubleshooting"
from page 109, you may not use the device.
– Have servicing and repairs carried out
exclusively by seca service or by an
authorized service partner.
– Please see the section entitled
"Servicing/recalibration" on page 111.
10.TROUBLESHOOTING
Problem
Cause / Solution
No weight displayed
with a load on the scale.
The scale has no power supply.
- Check whether the scale is switched on.
0.0 does not appear
before weighing.
A load was already on the scale before it was
switched on.
- Remove the load from the scale.
- Switch the scale off and back on again.
One segment is illumi-
nated either continu-
ously or not at all.
The corresponding point has a fault.
- Inform the service department.
The display
appears.
Battery block power is running down.
- Recharge battery block as soon as possible.
The display "bAtt"
appears.
Battery block discharged.
- Charge battery block.
The display "StOP"
appears.
Maximum load has been exceeded.
- Remove the load from the scale.
The display "tEMP"
appears.
The ambient temperature of the scale is too high
or too low.
- Set up the scale in an ambient temperature
between +10 °C and +40 °C.
- Wait for around 15 minutes until the device
has adapted to the ambient temperature.