What do i do if – Seca 762 User Manual
Page 39
What do I do if …? • 39
English
If you find faults or deviations during the function
check, first try to remedy the fault with the aid of the
section “What do I do if …?”.
CAUTION!
Bodily injury
If you find faults or deviations during the
function check which you are unable to remedy
with the aid of the section “What do I do if …?”,
you may not use the device.
– Have the device repaired by seca service or
an authorized service partner.
– You should also follow the instructions in the
section “Maintenance/Recalibration” on
page 40.
9. WHAT DO I DO IF …?
Malfunction
Cause/Remedy
... the scale does not
show any weight?
The device is not ready for use
- Remove transport lock
... the scale does not
show any plausible
weight?
• Before weighing pointer does not read “0”
- Adjust display
• Force shunt: scale is e.g. touching a wall or
patient is supporting his/her weight against the
wall.
- Move scale away from wall
- Use transport lock as wall spacer
- Ask patient to stand on the scale unaided