2 desinfection, 3 sterilisation, Function check – Seca 762 User Manual
Page 38

38 •
7.2
Desinfection
ATTENTION!
Damage to device
The viewing windows of scales and displays are
made of acrylic glass. Acrylic glass windows
can crack or become cloudy if unsuitable
disinfectants are used on them.
– Only use disinfectants suitable for sensitive
surfaces. Suitable disinfectants are available
from specialist dealers.
1. Follow the instructions on the disinfectant.
2. Disinfect the device at regular intervals using a soft
cloth dampened with a suitable disinfectant.
3. Please take note of the following terms:
7.3
Sterilisation
This device may not be sterilised.
8. FUNCTION CHECK
Perform functional check prior to each operation.
A complete functional check includes:
• Visual inspection for mechanical damage
• Visual inspection and function check of the display
elements
• Function check of all controls shown in the section
“Overview”
Interval
Component
Prior to every measurement
with direct skin contact
Weighing platform
After every measurement
with direct skin contact
Weighing platform
If required
Scale ring and viewing
window