beautypg.com

Cleaning, What do i do if – Seca 799 User Manual

Page 23

background image

Model 799

23

GB

• BMI between 25 and 30 (pre-obese):

The patient is slightly to moderately overweight. He should cut his weight if he is

already ill (e.g. diabetes, hypertension, gout, fat metabolism disorders).

• BMI over 30:

Weight reduction is essential. Metabolism, cardiovascular system and bones are all un-
der strain. A consistent diet, plenty of movement and behavioural training are all rec-
ommended. If in doubt, consult a specialist.

6. Cleaning

Clean the rubber surface and the housing as required using a domestic cleaning agent or

commercially-available disinfectant. Follow the manufacturer’s instructions.
Under no circumstances use abrasive or acid cleaners, white spirit, benzene or the like

for cleaning. Such substances can damage the high-quality surfaces.

7. What do I do if…

… no weight display comes on?
– Is the scale switched on?
– Check the batteries.
– Is the plug contact on the scale properly connected??

0.0

does not appear before the weighing operation?

– Press the Start key again – there must not be any load on the scale

----

appears instead of

0.0

before weighing?

– Press the Start key again – there must not be any load on the scale.

… one segment is illuminated constantly or not at all?
– The relevant segment has a fault. Inform Servicing.

… the display shows

?

– Battery voltage is running low. Change the batteries in the next few days.

bAtt

appears in the display?

– Batteries are empty. Put in new batteries.

St0p

appears in the display?

– Maximum load has been exceeded.

… the display flashes?
– If you have just activated the Hold or Tare function, please wait until the scale has de-

tected a steady weight value. The display will then stop flashing.

– If there has been no activation of a function beforehand, remove the load from the scale

and wait until

0.0