Fitting the column

The column and display head are fitted to the weighing platform.
– Guide the connecting cable through the

opening.

– Place the column in the recess. The

display should be legible for people

standing on the scale.

– Screw the column to the weighing plat-

form from underneath.

– If necessary, lay the scale carefully on

its side to tighten up the screws.

– Connect the plug of the connecting

cable to the rear of the scale.

– Then place the scale upright again.

4. Setting up and aligning the scale

– Place the scale on a level surface.
– Unfix the locknuts above the four foot

screws (1).

– Align the scale by screwing the foot

screws. The air bubble in the spirit level

must be right in the centre of the

circle (2).

– Then re-tighten the locknuts (3).
– Ensure that only the feet of the scale are

in contact with the floor. The scale may

not be in contact at any other point.

Important!

The alignment of the base of the scale

must be checked and corrected if nec-
essary every time the location of the

scale is changed.

1.

2.

3.

spirit level