Seca 757 User Manual
Page 81

Technical data • 81
English
Ambient conditions, storage
• Temperature
• Air pressure
• Humidity
-10 °C to +65 °C (14 °F to 149 °F)
700 - 1060 hPa
0 % - 95 %, no condensation
Ambient conditions, transport
• Temperature
• Air pressure
• Humidity
-10 °C to +65 °C (14 °F to 149 °F)
700 - 1060 hPa
0 % - 95 %, no condensation
Height of digits
11 mm
Power supply
Battery block, power pack
Power consumption
• with deactivated wireless module and
without background lighting
• with activated wireless module and per-
manent background lighting (100 %)
approx. 47 mA
approx. 100 mA
Maximum runtime in battery mode
• with deactivated wireless module and
without background lighting
• with wireless module activated
approx. 2880 minutes
Power pack recommended
Medical device in accordance with
Directive 93/42/EEC
Class I with measuring function
EN 60 601-1:
• Class II insulated appliance:
• Type B electrical medical device:
Type of protection
IP20
Duty cycle
Continuous duty
Wireless transmission
• Frequency band
• Transmission power
• Standards applied
2.433 GHz - 2.480 GHz
< 10 mW
EN 300 328
EN 301 489-1
EN 301 489-17
General technical data
seca 757