Starting weighing, Taring off additional weight (tare) – Seca 757 User Manual
Page 63

Operation • 63
English
Starting weighing
1. Ensure that there is no load on the scale.
2. Press the Start key.
All the elements of the display are shown briefly,
then SECA appears in the display.
The scale is ready for operation when the display
reads 0.000.
If the scale is being operated with a power pack,
the symbol
will appear in the display.
3. Place the baby on the tray.
4. Read off the measured result.
Taring off additional
weight (TARE)
You can use the TARE function to prevent the weighing
result from being affected by an additional weight (e.g.
a towel or pad on the tray).
NOTICE!
Incorrect measurement due to force shunts
If an additional weight (e.g. large towel) is in
contact with the surface on which the scale is
standing, the weight will not be measured
correctly.
– Make sure that any additional weight is only
placed on the tray of the scale.
1. Place the additional weight on the tray.
2. Hold down the (hold/tare) arrow key until the
message “NET” appears in the display.
3. Wait until the display stops flashing and is replaced
by 0.000.
4. Place the baby on the tray.
5. Read off the measured result.
The additional weight has been deducted
automatically.
6. To disable the TARE function, press the (hold/tare)
arrow key until the message “NET” is no longer dis-
played or switch off the scale.
NOTE
The maximum weight which can be displayed is
reduced by the weight of the objects already
placed on the scale.
hold
tare
hold
tare