11. Technical data
Dimensions
depth: 320 mm
width: 551 mm
height: 163 mm
Weight
approx. 6.5 kg
Temperature range
+10 °C to +40 °C
Approval as per
Directive
2009/23/EC
Class III medical
approval
Maximum capacity
Weighing range 1
15 kg
Weighing range 2
6 kg
Minimum capacity
Weighing range 1
0.1 kg
Weighing range 2
0.04 kg
Graduations
up to 6 kg:
2 g
up to 15 kg:
5 g
Accuracy at first calibration
Weighing range 1
0 up to 2.5 kg:
± 2.5 g
2.5 kg up to 10 kg: ± 5 g
10 kg up to 15 kg: ± 7.5 g
Weighing range 2
0 up to 1 kg:
± 1 g
1 kg up to 4 kg:
± 2 g
4 kg up to 6 kg:
± 3 g
EN 60 601-1:
insulated equipment,
protection class II,
electrical medical
equipment, Type B
Medical product
as per Directive
93/42/EEC
Class I
12. Spare parts and accessories
Spare parts
Plug-in mains unit
230 V~/50 Hz/12 V=/150 mA
seca no.: 68-32-10-252
Switchmode power adapter
100-240V~ / 50-60Hz / 12V= / 0.5A
seca no. 68-32-10-265
Accessories
seca 231 measuring rod for baby
scales
Measuring range:
350 – 800 mm
14 – 32"
Graduations:
1 mm
1/8"
Weight:
2 kg
Item no. 231 1717 009
seca 429 head and foot support for
baby scales
Item no. 429 0000 009
13. Disposal
Disposing of the device
Do not dispose of the device in
domestic waste. The device must
be disposed of properly as elec-
tronic waste. Follow the national regula-
tions which apply in your case.
For further information, contact our service
department at:
[email protected]