Yokogawa FU24F User Manual

Page 10

background image

IM 12B6J8-01E-E

10

CSA

Certificate no.

: 2516979

Master contract no. : 182892

IS, Class I Div. 1, GP A, B, C, D T3…T6

Electrical data

: For sensor input circuits (by connector), connected to a certified

intrinsically safe circuit, with the following maximum values

: Ui = 6.1 V; Ii = 230 mA; Pi = 1.2 W; Li = 4 μH; Ci = 30 μF

or

Certified intrinsically safe Yokogawa transmitter Model FLXA21

series.

Ambient temperature : T6 for Tamb. -40 °C to +60 °C

T5 for Tamb. -40 °C to +75 °C

T4 for Tamb. -40 °C to +110 °C

T3 for Tamb. -40 °C to +125 °C

Note: Intrinsically safe when connected as per Control Drawing FF1-K1226QV (see Fig 2)

Control Drawing CSA

The FU24F SENCOM

TM

sensor shall be

installed to a certified intrinsically safe circuit

meeting the entity parameters of the sensor

as shown in the table as maximum values,

or to a certified intrinsically safe Yokogawa

transmitter Model FLXA21 series.

When installing this equipment, follow the

manufacturer’s control drawing. Installation

should be in accordance with Canadian

Electrical Code, Part 1 or CEC, Part 1.

To prevent ignition of

flammable or combustible

atmospheres, disconnect

power before servicing or

read, understand and adhere

to the manufacturer’s live

maintenance procedures.

81 82 83 84 85 86 87

+ - G

Hazardous

location

Transmitter

WU11 cable

terminals 82-84,

86-87.

Maximum cable

length: 100 m.

Cable diameter:

5.8mm

Sensor

Ui

6.1 V

Ii

230 mA

Pi

1.2 W

Li

4 µH

Ci

30 µF

Maximum values

FU24F

Model:

NPT Dome shape

FSM Flat surface

IS, Class I Div. 1, GP A, B, C, D T3/T4/T5/T6

Ta 125 ºC/110 ºC/75 ºC/60 ºC

Fig 2: FF1-K1226QV Control Drawing CSA