Name plate
Tag plate for flameproof type
Tag plate for intrinsically safe type
Tag plate for intrinsically safe type for EJX90A
F0211r.ai
D
D
AFTER DE-ENERGIZING, DELAY 5 MINUTES
BEFORE OPENING.
WHEN THE AMBIENT TEMP. ≥ 65°C,
USE THE HEAT-RESISTING CABLES ≥ 90°C
No. KEMA 07ATEX0109
Ex d IIC T6...T4, Ex tD A21, IP6X
Enlcosure : IP66, IP67
TEMP. CLASS T6 T5 T4
MAX PROCESS TEMP.(Tp.) 85 100 120 °C
Tamb. -50(-15) to 75 80 75 °C
T85°C(Tamb.:40°C, Tp.:80°C),T100°C(Tamb.:60°C, Tp.:100°C),
T120°C(Tamb.:80°C, Tp.:120°C) Min.Tamb.:-40(-15) °C(for Dust)
WARNING
No. KEMA 04ATEX1116 X
EEx ia IIB/IIC T4 Tamb.:-40(-15) to 60°C
MAX PROCESS TEMP.(Tp.):120°C
T85°C(Tp.:80°C), T100°C(Tp.:100°C), T120°C(Tp.:120°C)
Enclosure: IP66 and IP67
FISCO Field device(IIC) FISCO Field device(IIB) Entity Parameters
Ui=17.5V
Ui=17.5V
Ui=24V
Ii=380mA
Ii=460mA
Ii=250mA
Pi=5.32W
Pi=5.32W
Pi=1.2W
Ci=1.76nF
Ci=1.76nF
Ci=1.76nF
Li=0µH
Li=0µH
Li=0µH
D
No. KEMA 06ATEX0278 X
EEx ia IIB/IIC T4 Tamb.:-40(-15) to 60ºC
MAX PROCESS TEMP.(Tp.):120ºC
T85ºC(Tp.:80ºC), T100ºC(Tp.:100ºC), T120ºC(Tp.:120ºC)
Enclosure: IP66 and IP67
FISCO Field device(IIC) FISCO Field device(IIB) Entity Parameters
Ui=17.5V
Ui=17.5V
Ui=24V
Ii=380mA
Ii=460mA
Ii=250mA
Pi=5.32W
Pi=5.32W
Pi=1.2W
Ci=1.76nF
Ci=1.76nF
Ci=1.76nF
Li=0µH
Li=0µH
Li=0µH
MODEL: Specified model code.
STYLE: Style code.
SUFFIX: Specified suffix code.
SUPPLY: Supply voltage.
OUTPUT: Output signal.
MWP: Maximum working pressure.
CAL RNG: Specified calibration range.
NO.: Serial number and year of production*
1
.
TOKYO 180-8750 JAPAN:
The manufacturer name and the address*
2
.
*1:
The first digit in the final three numbers of the serial
number appearing after “NO.” on the name plate indicates
the year of production. The following is an example of a
serial number for a product that was produced in 2010:
The year 2010
91K819857 032
*2:
“180-8750” is the Zip code for the following address.
2-9-32 Nakacho, Musashino-shi, Tokyo Japan
( II )
6.4.6 Units That Can Be Displayed on the
LCD by the Automatic Link Function
INDEX
Unit
Display
1000
K
Kelvin
1001
°C
degC
1002
°F
degF
1010*
m
m
1011*
km
km
1012*
cm
cm
1013*
mm
mm
1018*
ft
ft
1019*
in
in
1020*
yd
yd
1034*
m
3
m3
1035*
dm
3
dm3
1036*
cm
3
cm3
1037*
mm
3
mm3
1038*
L
L
1039*
cl
cl
1040*
ml
ml
1041*
hl
hl
1042*
in
3
in3
1043*
ft
3
ft3
1044*
yd
3
yd3
1045*
mile
3
mile3
1046*
pint
pint
1047*
quart
quart
1048*
gallon
gallon
1049*
ImpGal
ImpGal
1050*
bushel
bushel
1051*
bbl
bbl
1052*
bbl(liq)
bblliq
1088*
kg
kg
1089*
g
g
1090*
mg
mg
1091*
Mg
Mg
1092*
t
t
1094*
lb
lb
1097*
kg/cm
3
kg/cm3
1100*
g/cm
3
g/cm3
1101*
g/m
3
g/m3
1103*
kg/L
kg/L
1104*
g/ml
g/ml
1105*
g/L
g/L
1130
Pa
Pa
1131
GPa
GPa
1132
MPa
MPa
1133
kPa
kPa
1134
mPa
mPa
1135
µPa
uPa
1136
hPa
hPa
1137
bar