beautypg.com

Sidestream etco, Setup – ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual

Page 8

background image

E Series - End Tidal Carbon Dioxide (EtCO

2

) Option Insert

6

9650-1212-01 Rev. E

Applying an Airway Adapter with Mouthpiece

The disposable Pediatric/Adult airway adapter with
mouthpiece can be used for spot checking CO

2

on

non-intubated adult or pediatric patients.

CAUTION! The disposable Pediatric/Adult Airway

Adapter with mouthpiece is intended for
single patient use. Do NOT reuse or sterilize
the adapter, as system performance will be
compromised.

1. Remove adapter with mouthpiece from the package.

Verify that the adapter and mouthpiece are intact and
securely fastened to each other.

2. Attach the airway adapter to the CAPNOSTAT 5 CO

2

sensor, as follows:

a. Align the arrow on the bottom of the airway adapter

with the arrow on the bottom of the CAPNOSTAT
sensor.

b. Press the sensor and airway adapter together until

they click (see Figure 3).

3. If the unit displays the CHECK CO2 ADAPTER

message, follow steps a through c, else go to step 4.

a. Verify proper connection of the adapter to the

sensor.

b. Verify that the airway adapter windows are clean

and dry.

c. If the adapter is properly connected, and the

windows are clean and dry, then zero the adapter
as described in “Zeroing the Mainstream
CAPNOSTAT 5 CO2 Sensor/Airway Adapter” on
page 5.

4. Ensure the patient seals his or her mouth completely

around the mouthpiece and breathes normally.

A nose clip may be needed if the patient is exhaling
through the nose. It is important that all, or most, of the
exhalation be routed through the airway adapter.

Sidestream EtCO

2

Setup

There are several steps involved with sidestream EtCO

2

setup. These steps include:

Attaching the LoFlo Module Cable

Selecting a Sidestream Airway Adapter Kit or
Cannula

Inserting the Sample Cell

Zeroing the LoFlo CO

2

Module/Sample Cell

Applying a Sidestream Airway Adapter Kit

Applying a Nasal or Nasal/Oral Cannula

Attaching the LoFlo Module Cable

To attach the LoFlo module cable, plug the cable into the
yellow CO

2

connector at the back of the E Series unit by

matching the key on the cable to the key on the
connector.

To remove the sensor cable from the E Series unit,
grasp the collar surrounding the cable’s E Series
connector and pull up.

Selecting a Sidestream Airway Adapter Kit

Select an airway adapter kit based on the patient's size,
ET tube diameter, and monitoring situation. Airway
adapter kits are disposable and single patient use.

Note For monitoring times exceeding 6 hours, Nafion

tubing is recommended.

Figure 3

CO

2

Sensor

CAPNOSTAT

Mouthpiece

To Patient

Airway
Adapter

Airway Adapter Kit

ET Tube
Diameter

Adult/Pediatric Airway Adapter Kit

> 4.0 mm

Adult/Pediatric Airway Adapter Kit
with Nafion® tubing

Pediatric/Infant Airway Adapter Kit

≤ 4.0 mm

Pediatric/Infant Airway Adapter Kit
with Nafion tubing

Proper Grasp

Improper Grasp

Figure 4