beautypg.com

ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual

Page 22

background image

E Series - End Tidal Carbon Dioxide (EtCO

2

) Option Insert

20

9650-1212-01

Rev.

E

Software Hazards

Minimized by compliance with EN1441

Operating Time

EtCO

2

and SpO

2

Options

For a new fully charged PD4410 battery
pack at 20º C (68° F):

35 defibrillator discharges at maximum
energy (200 J), or

1.5 hours minimum of continuous ECG
monitoring, or

1.0 hour of continuous ECG monitoring/
pacing at 60 mA, 70 beats per minute.

For a new, fully charged lithium ion battery
pack at 20°C (68°F):

90 defibrillator discharges at maximum
energy (200 J), or

3.0 hours minimum of continuous ECG
monitoring, or

2.5 hours of continuous ECG
monitoring/pacing at 60 mA, 70 beats
per minute.

CAPNOSTAT 5® CO

2

Sensor

LoFlo™