Specifications – ZOLL E Series Monitor Defibrillator Rev E ETCO2 User Manual
Page 21
![background image](/manuals/355790/21/background.png)
9650-1212-01 Rev. E
19
Specifications
Specifications
This section summarizes the specifications of the E Series End Tidal Carbon Dioxide (EtCO
2
) option.
CAPNOSTAT 5® CO
2
Sensor
LoFlo™
Transducer Type
Mainstream
Sidestream
Principle of Operation
Non-Dispersive Infrared (NDIR) single beam optics, dual wavelength, no moving parts.
Warm Up Time
Full specifications within 2 minutes at an ambient temperature of 25º C. Capnogram in
15 seconds.
EtCO
2
Measurement Range
0 - 150 mmHg
0 - 20%
0 - 20 kPa
EtCO
2
Accuracy
(at 760 mmHg, ambient
temperature of 25°C)
0 - 40 mmHg, ±2 mmHg
41 - 70 mmHg, ±5% of actual
71 - 100 mmHg, ±8% of actual
101 - 150 mmHg, ±10% of reading
(At respiration rates > 80 breaths per minute, all ranges are ±12% of actual.)
EtCO
2
Resolution
0.1 mmHg 0 - 69 mmHg
0.25 mmHg 70 - 150 mmHg
EtCO
2
Stability
Short-Term Drift: Drift over four hours < 0.8 mmHg.
Long-Term Drift: Accuracy specification will be maintained over a 120 hour period after
zeroing.
EtCO
2
Noise
RMS noise of the sensor < 0.25 mmHg at
7.5% CO
2
.
RMS noise of the sensor < 0.25 mmHg at
5% CO
2
.
EtCO
2
Rise Time (10-90%)
< 60 ms (Adult/pediatric adapters)
< 60 ms (Infant/pediatric adapters)
< 3 seconds (includes transport and rise
time)
Respiration Rate Range
2 - 150 breaths per minute
Respiration Rate Accuracy
±1 breath per minute
Sample Flow Rate
N/A
50 ml/min ±10 ml/min
Compensations
Barometric pressure 400 - 850 mmHg (automatic).
Operator selectable O
2
/N
2
O compensation.
EtCO
2
Alarm Limits
User selectable, High 5 - 100 mmHg, Low 0 - 95 mmHg, OFF.
Respiration Rate (RR) Alarm
Limits
User selectable, High 5 - 150 respirations per minute, Low 0 - 100 respirations per
minute, OFF.
Halogenated Agents
Specification allows for halogenated anesthetic agents that may be present at normal
clinical levels. The presence of Desflurane in the exhaled breath beyond normal values
(5-6%) may positively bias carbon dioxide values by up to an additional 2-3 mmHg.
Airway Adapter Deadspace
Adult < 5 cc
Infant < 1.0 cc
Adult < 7.3 cc Maximum
Pediatric/Infant < 1.0 cc
Environmental (sensor or
module)
Operating Temperature: 0º C to 45º C (0º C to 40º C for LoFlo Module)
Storage and Shipping Temperature: -40º C to 70º C
The E Series unit may not perform to specifications when stored at the upper or lower
extreme limits of storage temperature and immediately put into use.
Electromagnetic Immunity
AAMI DF-80:2003, IEC 60601-1-2, 10 V/m