VBT-60/80 Operating Procedures

11

Table 3.0. Functional Description of VBT-60 Controls and Display

Figure 1

Index #

Adjacent Panel Marking

Functional Description

1

(Connector)

High-voltage-cable connector

2

RS-232C

RS-232C interface port; 9-pin connector;
female DB type. The data are set to 19,200
baud, 1 start bit, 8 data bits, and no parity bit;

PIN ...............SIGNAL

2 Rx
3 Tx
5 Signal Gnd
This serial port is for factory calibration and
firmware updates

3

90-230 Vac, 8A, 50- 60 Hz

Input power connector with built-in fuse
holder and power switch.

4

GROUND

(Wing Nut)

VBT ground stud. Connect ground stud to
substation ground using the provided cable.

5

No marking

LCD; 2-line by 16-character; back-lighted;
displays menus of selections, operator entries,
and test-measurement results.

6

TEST FAIL

Test Fail Indicator. This indicator is turned on
if the test current exceeds the preset current
threshold (100, 200, or 300

?

A).

7

CHANGE

SELECT

VBT control knob. Turn knob to desired new
feature. Press knob to select that new feature

8

HIGH VOLTAGE

ENABLE

LED indicator, red; Lights when high-test-
voltage is present at the test leads.

9

PUSH TO

“ARM”

Arm switch; Press and hold during test

10

GROUND

High-voltage-return cable connector.