beautypg.com

0 dmom-100, Specifications, Table 1.0 dmom-100 – Vanguard DMOM-100 User Manual

Page 8: Specifications and leading particulars, Operating procedures dmom-100

background image

Operating Procedures

DMOM-100™

2

2.0 DMOM-100

SPECIFICATIONS

Table 1.0

DMOM-100

Specifications and Leading Particulars

MODEL .............................................DMOM-100

 (part number VIC-9000-100)

TYPE..……… ..................................Special-purpose test equipment, portable; 100-ampere,

Digital Micro-Ohm Meter

POWER ............................................85 to 132 V ac or 200-240 V ac (selectable), 50/60 Hz

SIZE (inches) ....................................16.8 wide by 12.6 high by 10.6 deep

WEIGHT............................................Less than 21 lbs

TEST CURRENT ..............................10 to 100 amperes

RESISTANCE READING RANGE ...1 micro-ohm to 300 milliohms

ACCURACY .....................................

±1% reading, ± 1 count

DISPLAY ..........................................LCD, 4-line by 20-character, back lighted

CONTROLS......................................Keypad: 10 number keys and 6 function keys

COMPUTER INTERFACE ................RS-232C

PRINTER ..........................................Thermal printer (prints on 2.5-inch-wide thermal paper)

ENVIRONMENT ...............................Operating: 0

°C to 55°C; Storage: -40°C to 65°C

FURNISHED .....................................One power cord, one 35-foot test cable set,
One cable bag

EXPENDABLES ...............................Paper, thermally sensitive, 2.5-inch wide roll

(VIC part # TP-3)

WARRANTY .....................................One-year warranty on parts and labor; post warranty

service contracts available

NOTE:

THE ABOVE SPECIFICATIONS ARE VALID AT NOMINAL OPERATING VOLTAGE AND AT A TEMPERATURE OF 25

°C

(77

°F)

DMOM-100