beautypg.com

Teledyne LeCroy SDA II User Manual

Page 15

background image

Operator's Manual

PCI-E

XPRESS

There are four PCI-Express selections that meet the following standards:

l

PCI-Express Gen1

l

PCI-Express Gen2 A

l

PCI-Express Gen2 B

l

PCI-Express Gen2 C

DVI

The DVI selection displays the DVI PLL dialog at the far right of the screen and follows the requirements
of the DVI (Digital Video Interactive) and HDMI (High Definition Multimedia Interface) specifications.
These specifications call out a clock recovery function having a single-pole PLL loop response with a cutoff
of 4 MHz.

FB-DIMM
The FB-DIMM selection applies a filter that replicates Fully-buffered DIMM.

USB
The USB selection applies a filter that meets the USB 3.0 SS requirements.

C

USTOM

The Custom selection shows the Custom PLL dialog on the far right of the screen and allows you to select
either a first or second order loop response. The first order response allows you to select a pole
frequency that sets the PLL cutoff, and a zero frequency that must be higher than the pole frequency
that limits the stop-band attenuation.

The second order PLL allows you to select the natural frequency and damping factor. The damping factor
determines the transient behavior of the phase locked loop and is variable from 2 to 0.5.

l

A damping factor above 0.707 results in an under-damped response and the PLL is over-corrected
to a sudden change in frequency; however, reacts quicker to the change.

922968 Rev A

13