QPHY-MOST150o Software Option
920456 Rev B
5
INTRODUCTION TO QUALIPHY MOST150 OPHY
QPHY-MOST150o is an automated test package performing all of the normative real time oscilloscope
tests for sources in accordance with the MOST150 oPHY Automotive Physical Layer Sub-Specification
Rev. 1.1, 05/2010. The software can be run on LeCroy Zi and Xi Series oscilloscope with at least 2
GHz bandwidth, a sample rate of at least 10GSa/s and 10Mpt of memory. The oscilloscope must also
be equipped with the QPHY-MOST150o, DFP2, and either SDM or SDA2 options.
Required Equipment
• QPHY-MOST150o, DFP2, and either SDM or SDA2 options
• Real time LeCroy Oscilloscope meeting the above requirements (at least 2 GHz bandwidth, a
sample rate of at least 10GSa/s, and 10Mpt of memory)
o
All LabMaster 9 Zi-A and SDA/DDA/WaveMaster 8 Zi-A oscilloscopes
o
SDA/DDA/WavePro 725Zi-A and higher bandwidth oscilloscopes
o
WaveRunner 620 Zi and higher bandwidth oscilloscopes
o
WaveRunner 204 MXi-A/Xi-A Series
Note: If a WaveRunner Xi oscilloscope is being used, the XWEB option is additionally
required.
• LeCroy OE425 Optical to Electrical Converter or equivalent (for optical measurements)
• LeCroy D420 Differential Probe and WL-PLINK (for electrical measurements)
Note: If non-compliance testing is being performed, a pair of matched SMA cables can be
used in place of the differential probe. This may require SMA/BNC connectors on some
oscilloscope models.