QPHY-MOST150o Software Option

920456 Rev B

21

Additional tests performed at SP1

Check Bitrate 1.1

The actual bitrate of the signal is measured and the deviation from the nominal value is compared to
the specification.

Figure 14 - Check Bitrate Results

Additional tests performed at SP2

Dark level and Extinction Ratio 2.1

The dark level of the O/E converter is measured and used to calculate the extinction ratio. To
measure the dark level, remove the POF-fibre from the O/E-converter. Shade the input of the O/E-
converter and press the “OK” button. Skipping the measurement of the dark level is allowed but will
result in a failure of the extinction ratio test.

Figure 15 - Dark Level and Extinction Ratio Results