Teledyne LeCroy WaveLink Differential Probe Series (4, 6 GHz) User Manual
Page 68

62
923175-00 Rev A
Description
Minimum Requirements
Example Test Equipment
Oscilloscope, High BW
1
BW
³
6 GHz
Teledyne LeCroy: WaveMaster 8600A,
WavePro 760 Zi, WaveMaster 806 Zi/Zi-A
Oscilloscope, High BW
2
BW
³
4 GHz
Teledyne LeCroy: WaveMaster 804Zi-A,
WavePro 740 Zi, WaveRunner 640 Zi
Oscilloscope, High Impedance
200 mV/div - 2 V/div scale
factor 1 MΩ input impedance
ProBus interface
Teledyne LeCroy: WavePro 7300 or
WaveRunner 6200, WavePro 735 Zi-A,
WaveRunner 6 Zi
Digital Multimeter
AC: 0.2% accuracy
to
measure 200 mV and 2 V
rms
@
1 kHz 6½ digit resolution
Agilent Technologies: 34401A, or Fluke:
8842A-09, or Keithley: 2001
Oscillator/Function Generator
Sine Wave output, adjustable
from 500 mV to 4 Vp-p (357
mV to 2.83 V
rms
) at 70 Hz
Stanford Research: Model DS340,or
Agilent Technologies: 33120A, or Leader:
LAG-120B
Pulse Generator
12 ps, -5 V
out
, 2.4 mm output
Picosecond Pulse Labs: 4015D-215
Calibration Fixture
3
See Preliminary Procedure
Teledyne LeCroy: ProLink-CF01
Calibration Fixture
4
See Preliminary Procedure
Teledyne LeCroy: ProBus-CF01
Terminator, Precision, BNC
50 Ω ± 0.05%
Teledyne LeCroy: TERM-CF01
Characterization Fixture
5
Teledyne LeCroy: PCF-200
SMA to BNC Adapter
Female SMA to male BNC
Pomona Electronics: 4289