Appendix a, Performance verification test record, Equipment used – Teledyne LeCroy CP031 User Manual
Page 26: Cp031 test record
![background image](/manuals/353485/26/background.png)
CP031 Current Probe
20
923362-00 Rev A
Appendix A
Performance Verification Test Record
Photocopy this page and record the results of measurements made during the performance verification
of the CP031 Current Probe on the copy. File the completed record as required by applicable internal
quality procedures.
The sections in the test record correspond to the parameters tested in the performance verification
procedure. The numbers preceding the individual data records correspond to the steps in the procedure
that require the recording of data. Results recorded in the column labeled "Test Result" are the actual
specification limit check. The test limits are included in all of these steps. Other measurements and the
results of intermediate calculations that support the limit check are to be recorded in the column
labeled "Intermediate Results".
Permission is granted to reproduce these pages for the purpose of recording test results.
Model:
_______________________
Serial Number:
_______________________
Asset or Tracking Number:
_______________________
Date:
_______________________
Technician:
_______________________
Equipment Used
MODEL
SERIAL NUMBER
CALIBRATION DUE DATE
Digital Multimeter #1
Digital Multimeter #2
Function Generator
1
N/A
1
The function generator used in this Performance Verification Procedure is used for making relative
measurements. The output of the generator is measured with a DMM or oscilloscope in this procedure.
Thus, the generator is not required to be calibrated.
CP031 Test Record
Step
Description
Test Result
Gain Accuracy
10 Probe Output (Spec limit: 0.495 - 0.505 V)
________________ V