background image

Instructions

Deskewing Master and Slave Channels

Deskewing is a procedure primarily used to synchronize the timing of the waveforms displayed on screen
in order to account for propagation delays through probes and cables, and transfer times from the Slave
to the Master oscilloscope.

When you select Master + Slave control, the WR6Zi-8CH-SYNCH software automatically accounts for the
propagation delay of the supplied LBUS cable.  When making other types of connections between the
instruments, each of the Slave and Master channels can be fine adjusted for skew.

Both the WaveRuner 6 Zi and the HRO 6 Zi scopes have a fast edge output using the front panel Aux Out-
put port. This fast edge represents the common signal from which to adjust the deskew values.

Warranty

THE WARRANTY BELOW REPLACES ALL OTHER WARRANTIES, EXPRESSED OR IMPLIED, INCLUDING BUT NOT LIM-
ITED TO ANY IMPLIED WARRANTY OF MERCHANTABILITY, FITNESS, OR ADEQUACY FOR ANY PARTICULAR PURPOSE
OR USE. TELEDYNE LECROY SHALL NOT BE LIABLE FOR ANY SPECIAL, INCIDENTAL, OR CONSEQUENTIAL DAMAGES,
WHETHER IN CONTRACT OR OTHERWISE. THE CUSTOMER IS RESPONSIBLE FOR THE TRANSPORTATION AND INSUR-
ANCE CHARGES FOR THE RETURN OF PRODUCTS TO THE SERVICE FACILITY. TELEDYNE LECROY WILL RETURN ALL
PRODUCTS UNDER WARRANTY WITH TRANSPORT PREPAID.

The product is warranted for normal use and operation, within specifications, for a period of three years
from shipment. Teledyne LeCroy will either repair or, at our option, replace any product returned to one
of our authorized service centers within this period. However, in order to do this we must first examine
the product and find that it is defective due to workmanship or materials and not due to misuse, neglect,
accident, or abnormal conditions or operation.

Teledyne LeCroy shall not be responsible for any defect, damage, or failure caused by any of the fol-
lowing: a) attempted repairs or installations by personnel other than Teledyne LeCroy representatives or
b) improper connection to incompatible equipment, or c) for any damage or malfunction caused by the
use of non-Teledyne LeCroy supplies. Furthermore, Teledyne LeCroy shall not be obligated to service a
product that has been modified or integrated where the modification or integration increases the task
duration or difficulty of servicing the oscilloscope. Spare and replacement parts, and repairs, all have a
90-day warranty.

Products not made by Teledyne LeCroy are covered solely by the warranty of the original equipment
manufacturer.