Operation – Teledyne LeCroy WaveRunner 6 Zi_HRO 6 Zi, 8-Channel Sync Instruction Manual User Manual
Page 3

Instructions
Operation
Connecting Slave to Master
Follow these steps to open the communication link between the Master and Slave instruments.
1. From the menu bar of the Master oscilloscope, choose Vertical → Remote Slave DSO to display the
Remote Slave DSO dialog.
2. Confirm that the L-Bus cable Connected box is checked, signaling that the cable is correctly con-
nected to both instruments.
3. From the menu bar of the Slave oscilloscope, choose Utilities Setup → Remote. Note the Slave IP
Address shown.
4. In the Slave IP Address field on the Master Remote Slave DSO dialog, enter either the Slave IP
Address or oscilloscope name (host name).
l
The Master will display a brief message in the lower left-hand part of the display confirming suc-
cessful connection to the Slave.
l
The Connect Slave DSO control will be disabled, and Disconnect Slave DSO will be enabled.
5. Confirm Slave Connected box is checked.
Disconnecting Slave from Master
Follow this disconnection procedure to close the communication and waveform transfer between
Master and Slave instruments.
1. On the Master oscilloscope, choose Vertical → Remote Slave DSO.
2. Click the Master Only button. This will return the Slave timebase reference clock source to Internal.
3. Click the Disconnect Slave DSO button. This closes the Ethernet communication link.
4. Optionally, disconnect the Ethernet cable and LBUS cable connecting the two instruments.
NOTE: You can reestablish communication between Master and Slave by clicking Connect Slave DSO.
Waveform transfer and display from Slave to Master is reestablished by clicking Master + Slave.