Performance Profi le
Flow Range:
10 to 30 L/min
Flow Compensation
Accuracy:
± 5% of set fl ow
Compensating Flow
10 L/min at 90 inches water back pressure
Back Pressure Range: 20 L/min at 50 inches water back pressure
30 L/min at 15 inches water back pressure
Power: •
Rechargeable lithium-ion (Li-Ion) battery, 7.4 V, 9.6-Ah
capacity, 71.04 Wh
•
Battery with AC charger/adapter, 100-240 V
Battery Recharge Time: Approximately 5 hrs
(varies with batt ery capacity
and level of discharge)
Typical Cumulative
Spore Trap* (e.g., VersaTrap): 5 hrs at 15 L/min (battery only)
Run Time:
†
BioStage viable cascade impactor: 4 hrs at 28.3 L/min (battery only)
25-mm, 1.2-μm MCE fi lter: 9+ hrs at 10 L/min
To achieve, use QuickTake 30 with a fully
charged battery and AC charger/adapter.
37-mm, 0.8-μm MCE fi lter: > 14 hrs at 10 L/min (battery only)
* Sampling times when using spore traps are usually ≤ 10 min. SKC
recommends reducing length of sample time when using a 30 L/min
fl ow rate to prevent overloading the media.
† Results obtained using a new pump and new fully charged batt eries. Pump and
batt ery performance may vary.
The AC charger/adapter can be used to extend batt ery run
time, but it is not a batt ery eliminator. Therefore, it will
not provide indefi nite run times.
Temperature:
Operating:
32 to 113 F (0 to 45 C)
Storage:
-4 to 95 F (-20 to 35 C)
Charging:
32 to 113 F (0 to 45 C)
Operating Humidity:
0 to 95% non-condensing
Protect sample pump from weather when in use outdoors.
Compare the information in this table to pump compensation range to determine appropriate applications.
Typical Back Pressure of Sampling Media (inches water)
Flow Rate (L/min)
10
12
15
20
Filter/Pore Size (μm)
25-mm MCE, 0.8
65
80
103
148
25-mm MCE, 0.45
138
172
> 200
—
37-mm MCE, 0.8
22
28
36
51
37-mm PVC, 5.0
9
11
15
21