Calibration and setting flow – SKC 228-9530 QuickTake 30 User Manual
Page 14

13
Tubing
Calibrator outlet
Pump inlet
Spore trap cassette
Note: Use a second cassette adapter to
attach calibrator to a spore trap cassette.
Digital display:
Calibration mode
Calibration with a spore trap cassette
Cassette adapter
Digital display:
Store setting
Calibration and Setting Flow
Use a primary standard calibrator to calibrate pump fl ow rate.
Calibrate the fl ow rate with a representative sampling medium in line.
Before use, allow the pump to equilibrate aft er moving it from one temperature
extreme to another.
1. Ensure the pump has run for 5 minutes before calibrating. Using 3/8-inch
ID fl exible tubing and appropriate adapters, connect the inlet of the pump
to the outlet of a representative sampling medium and a primary standard
calibrator to the inlet of the representative sampling medium (as shown
above).
2. Scroll to any run time preset. Press and hold the CAL
butt on for two seconds. The pump will start running
and a fl ashing CAL will appear on the LED.
3. Press or to increase or decrease the fl ow until
the desired fl ow rate is displayed on the calibrator.
4. Press
. A fl ashing Stor will appear on the LED.
To save new sett ing: Press [
]. The pump
will stop running.
To ignore new sett ing: Press
. The pump will
stop running.
Note:
A security code is not needed to change fl ow
rate. It is only required when changing factory
sett ings. To change factory sett ings, see Advanced
Setup.