Advanced setup, Intermittent sampling – SKC 228-9530 QuickTake 30 User Manual
Page 19

18
Advanced Setup
Intermittent Sampling
Intermitt ent Sampling allows the pump to be programmed to run for a specifi c
number of minutes up to 999, to shut off for a programmed
length of time, and to continue sampling on and off for a
predetermined number of cycles (e.g., program the pump
to run three cycles (n) of 15 minutes each (r) with a fi ve-
minute delay (d) between each cycle).
Entering Intermittent Sampling Mode
Press or to scroll to Intt . Press
to enter the intermitt ent sampling setup mode.
Setting Run Time (r)
1. Press or to increase or decrease run time. The display will fl ash. If no
change to run time is desired, press
to move to delay time.
2. When fi nished, press
. Stor will fl ash on the display.
To save the new sett ing: Press [
]. Press
to move
to delay time setup.
To ignore the new sett ing: Press
. Display will move
to the delay time.
If displayed values are not changed, the fl ashing Stor
will not appear and pressing
will move the display
to the next parameter to be set.
Digital display:
Intermittent Sampling
Digital display:
Set run time
Setting Delay Time (d)
1. Press or to increase or decrease delay time. The
display will fl ash. If no change to delay time is desired,
press
to move to number of cycles.
2. When fi nished, press
. Stor will fl ash on the display.
To save the new sett ing: Press [
]. Press
to move
to number of cycles setup.
To ignore the new sett ing: Press
. Display will move to the number of cycles.
Digital display:
Set delay time