beautypg.com

Calibration, Sampling, Operation – SKC 225-367 Abrasive Blasting Sampler User Manual

Page 2

background image

Calibration

1. Ensure pump had run for 5 minutes before calibrating.
2. Using fl exible tubing, connect the outlet of the
Butt on Sampler to the inlet of a sampling pump.
3. Place the calibration adapter outlet onto the Butt on

Sampler inlet, and connect the calibration adapter
inlet to a primary standard calibrator.

4. Calibrate to 4 L/min following the instructions in the

pump and calibrator operating instructions.

5. Disconnect the calibrator and calibration adapter.

Replace the fi lter used to set the fl ow with a fresh
conditioned and weighed fi lter for sample collection.

Note: If using the Butt on Sampler with protective shield for abrasive blasting
sampling, calibrate the Butt on Sampler without shield using the Calibration
Adapter. Following calibration, push shield onto sampler inlet and sample.

Sampling

1. Ensure the fl ow rate has been calibrated (see Calibration).
2. Using fl exible tubing, connect the outlet of the Butt on Sampler to the inlet of

a sampling pump such as an AirChek

®

XR5000 Sampler.

3. For personal sampling, clip the Butt on Sampler onto a worker’s collar or

pocket near the breathing zone. Clip the pump onto the worker’s belt or

place it in a protective pouch. Start the pump and record the start time,

worker location, and fl ow rate.

Button Sampler
with Calibration
Adapter

Cap

Inlet
Section

PTFE
O-ring

Filter

Support
Screen

BUNA-N
O-ring

Body

Collar Clip

Outlet Plug

Operation

Wear gloves and use forceps when handling fi lters.

1. For gravimetric sampling, condition and weigh a 25-mm fi lter according to

method used.

2. Unscrew the inlet of the sampler counterclockwise until it comes off and

remove the PTFE O-ring.

3. Place the 25-mm fi lter on top of the support screen inside the sampler.

Note: Representative fi lters used for calibration do not require conditioning and
weighing.

4. Place the PTFE O-ring on top of the fi lter.
5. Replace the inlet and very gently turn clockwise until moderately tight.

Caution: Excessive tightening of the inlet section can twist and tear the fi lter.

;;;;;

;;;;;

;;;;;

;;;;;

;;;;;

;;;;;

This manual is related to the following products: