Airlite quick start guide – SKC 110-100 AirLite User Manual

Page 3

background image

1

AirLite Quick Start Guide

On/Off

1.

Lift control cover on top of pump.

2. Press On/Off butt on.

Setting Flow Rate (1000 to 3000 ml/min)

1. Set up sampling train.

2. Connect calibrator to inlet of sample medium.

3. Turn on pump and adjust fl ow adjustment screw on pump until

desired

fl ow rate is indicated on calibrator. Take a minimum of

three readings. Calculate fl ow as the average of three readings.

LED Indicators

Slow fl ash = normal run

Fast fl ash = fl ow fault

Double fl ash = low batt ery, approximate minimum one-hour run time

remaining

Flash every eight seconds = pump shutdown due to exhausted batt eries

Four fl ashes every six seconds = pump shutdown due to fl ow fault

Note: Indication of pump shutdown due to exhausted batteries will override all
other indicators.

AirLite uses three replaceable standard AA alkaline batt eries.