Airlite quick start guide – SKC 110-100 AirLite User Manual
Page 3
1
AirLite Quick Start Guide
On/Off
1.
Lift control cover on top of pump.
2. Press On/Off butt on.
Setting Flow Rate (1000 to 3000 ml/min)
1. Set up sampling train.
2. Connect calibrator to inlet of sample medium.
3. Turn on pump and adjust fl ow adjustment screw on pump until
desired
fl ow rate is indicated on calibrator. Take a minimum of
three readings. Calculate fl ow as the average of three readings.
LED Indicators
Slow fl ash = normal run
Fast fl ash = fl ow fault
Double fl ash = low batt ery, approximate minimum one-hour run time
remaining
Flash every eight seconds = pump shutdown due to exhausted batt eries
Four fl ashes every six seconds = pump shutdown due to fl ow fault
Note: Indication of pump shutdown due to exhausted batteries will override all
other indicators.
AirLite uses three replaceable standard AA alkaline batt eries.