beautypg.com

SecurityMan SVR2001 User Manual

Page 29

background image

28

ways: "TIMER ON" (CLOSED CIRCUIT ) ," TIMER OFF" (OPEN CIRCUIT) ,and

"IDLE" (no response) For example, if the output device is a light or

a siren, choose "TIMER ON.". So, when "Input 1" triggers, the circuit

relay of the output dry contact will become "CLDSED CIRCUIT", that means

the light will be turn on or the siren will be turned on ; choose "TIMER

OFF". so when "Input 1"triggers, the circuit relay of the output dry

contact will become "OPEN CIRCUIT" that means the light will be turn

off; or choose "IDLE" for no response, Either selection of "TIMER ON"

or "TIMER OFF" will guide you to the next page of the "OUTPUT" action

time length set-up.

The "TIMER ON" or "TIMER OFF" is in accord with the setting of O/P PRESET;

" N.C." in accord with " TIME ON," " N.O." in accord with "TIME OFF"

2. TRIGGER

This is to set the triggering conditions of the input terminal. There

are three settings in the "TRIGGER" sub-menu, "CONDITION", "DETECT"

and "STATUS".

CONDITION: Set the input terminal to be triggered at the "N.O." (OPEN

CIRCUIT) or at "N.C." (CLOSED CIRCUIT) relay condition. When the input

terminal detects OPEN CIRCUIT or CLOSED CIRCUIT it will trigger the

alarm. For example, if the input device is a motion sensor (which is

normally open), choose "N.C." for triggering circuit relay condition.

DETECT: This is to enable (ON) or disable (OFF) the terminal detection

function.

STATUS: This is to set the trigger for one time (ONCE) or continuously

(CONTINUE).