Gentec-EO QE and XLE4 User Manual
Page 35
![background image](/manuals/347656/35/background.png)
QE Series Joulemeter Instruction Manual Version 3.0
April 2014
35
and position on the detector).
Step 5: Repeat the first measurement (Step 3) to make sure that nothing changed during the procedure
to invalidate the calibration. A change larger than the uncertainty of your measurements means that
something in the laser or environment changed. You can add this to your ± uncertainty when you use the
attenuator or try to stabilize the laser and environment and begin again with Step 3.
The correction multiplier for the MAESTRO and an Oscilloscope will be given by:
attenuator
with
Reading
attenuator
without
Reading
f
T
(No unit)
Now use this calibration factor in the correction menu for the “Attenuator/Diffuser” when using it at the
wavelength established in Step 1.