Specifications for the qe8-b mt and bl – Gentec-EO QE and XLE4 User Manual
Page 20
QE Series Joulemeter Instruction Manual Version 3.0
April 2014
20
1.3.3 Specifications for the QE8-B MT and BL
Footnotes
Model
QE8SP-B-MT-D0
QE8SP-B-BL-D0/DA
Optical Absorber
MT
BL
Spectral Range
1, 7
0.19
– 20 µm
Calibrated Spectral Range
1, 7
0.248
– 2.5 µm
Typical Sensitivity
2, 6
2400 V/J
900 V/J
Calibration Uncertainty
6, 2, 7, 8, 10
± 4%
Repeatability
< 0.5 %
Max. Pulse Energy @ 1.064 µm
M-Link
S-Link-2
Maestro
3, 11
1.3 mJ
1.1 mJ
0.93 mJ
3.6 mJ
2.9 mJ
2.5 mJ
Noise Equivalent Energy (Typ. NEE)
M-Link
S-Link-2
Maestro
2, 10
50 nJ
50 nJ
80 nJ
100 nJ
100 nJ
150 nJ
Max. Repetition
Rate
2, 4, 10
1000 Hz
400 Hz
Typical Rise Time
(0-100%)
2, 10
30 µsec
30 µsec
Max. Pulse Width (Typ)
2, 5, 9, 10
10 µsec
10 µsec
Max. Energy Density
50 mJ/cm
2
@ 1.064µm, 7ns, 10 Hz
Max. Average Power Density
1W/cm
2
Max. Average Power
0.5 W
Dimensions
Ø38.1 x 27.4D mm
Weight
91 g
Aperture Size
7.8 x 7.8 mm
Aperture Area Size
0.608 cm
2
FOOTNOTES FOR QE-B SERIES SPECIFICATIONS: