Specifications for xle4 – Gentec-EO QE and XLE4 User Manual
Page 23
QE Series Joulemeter Instruction Manual Version 3.0
April 2014
23
1.3.5 Specifications for XLE4
Footnotes
Model
XLE4
Optical Absorber
XT : Metallic
Spectral Range
6
0.19
– 20 µm
Calibrated Spectral Range
6
0.35
– 2.5 µm
Typical Sensitivity
1, 5
1100 V/J
Calibration Uncertainty
1, 5, 6, 7, 9
± 4% at 1064nm
± 9% for other wavelengths
Repeatability
< 0.5 %
Max. Pulse Energy 1.064 µm
2
4 mJ
Noise Equivalent Energy (NEE)
(Typ)
1, 9
150 nJ
Max. Repetition
Rate
1, 3, 9
2000 Hz
Typical Rise Time
(0-100%)
1, 9
10 µsec
Max. Pulse Width (Typ)
1, 4, 8, 9
5 µsec
Max. Energy Density
90 mJ/cm
2
@ 1.064µm, 7ns, 10 Hz
Max. Average Power
0.4 W
Dimensions (H x W x D)
26.5 x 36.0 mm diam
Weight
130 g
Aperture Size
4.0 mm diameter
Aperture Area Size :
0.16 cm
2
FOOTNOTES FOR XLE4 SERIES SPECIFICATIONS:
1
Load capacitance must be
13 pF, excluding the supplied BNC to DB-
15 “Smart Interface” coaxial
cable.
190 Assuming 90 mJ/cm
2
@ 1.064µm,7ns laser beam; with a uniform energy distribution; energy
applied to full aperture. Increasing the pulse width increases the maximum measurable energy.
190 At constant power.
4
To maintain indicated calibration, sensitivity must be de-rated for longer pulses.
5
Calibrated at: 400 µJ, 10 Hz, @1.064 µm, 7 ns pulse (FWHM), semi-Gaussian beam profile, energy
applied to 80% of aperture, loaded into 1 M
/ 13 pfd.
6
For values other than calibration wavelength a typical value is recommended but not traceable to NIST.
7
Excludes non-linearities.
8
Duration at base of pulse. Divide by 2 for FWHM (Full Width at Half Maximum) duration.
9
Loaded into
1 M
/ 13 pfd.