Conductivity

1. Make sure the sensor is Conductivity cell, or check that the icon

,

or

appears in the LCD display.

2. Dip the conductivity cell into the standard solution 1413

µS/cm. Stir gently and wait until the

display is stabilized. Press and hold

button to enter calibration mode until the icon

CAL

appears in the display, and flashes 1413

µS/cm. When the display stops flashing and

indicates “SA”, it will be followed by “End” while calibration is ending, and then return to
measurement mode.

3. If reading is not 0

µS/cm while the meter is in the air and is not dipped it into any solution,

then calibrate it in the air to make reading becomes 0

µS/cm.

Note: (1) Calibration with a 12.88 mS/cm standard solution would be better for measuring high

conductivity solution.

(2) The icon

will appear automatically during calibration mode.

(3) “SA” will not appear if the calibration fails.

MEASUREMENT
pH

1. After calibration, rinse the electrode with clean water and wipe it dry. Dip the electrode into

sample solution to be measured. Stir gently and wait until a stable reading can be obtained.

ORP

1. Insert ORP electrode and the icon

will appear in the LCD display automatically.

2. Calibration is not necessary for ORP. But it could be tested with a specific ORP solution to

check if the electrode is good or bad.

3. Rinse the electrode with clean water and wipe it dry. Dip the electrode into sample solution

to be measured. Stir gently and wait until a stable reading can be obtained.

Note: (1) “_ _ _ _” will appear in the display when it is over the measuring range.

(2) After measurement, rinse the electrode with clean water. Place the protective cap and

soaking bottle back on. The soaking bottle should be always kept wet by adding
soaking solution.

Conductivity >

1. After calibration, rinse the conductivity cell with clean water and wipe it dry. Dip the

conductivity cell into sample solution to be measured. Stir gently and wait until a stable
reading can be obtained.

2. Press

button to select the desired measurement mode Conductivity, TDS or Salinity.

Note: (1) “_ _ _ _” will appear in the display when it is over the measuring range.

(2) The unit will auto-range to

µS/cm or mS/cm, ppm or ppt.

(3) After measurement, rinse the electrode with clean water and place the protective cap

back.

TDS

Salt

COND

ORP

COND